Six T-Marker capture mistakes and how to avoid them

Most failed T-Marker captures come from a handful of setup mistakes. The manual shows each of them with a "Do" and "Avoid" example.

1. Markers facing different directions

Position every marker facing forward. Markers turned in different directions are harder for the camera to read.

2. Covered markers

Nothing should sit over a marker: no cheek retractor edge, no gauze, no finger. All dots and the code must be visible or accuracy suffers.

3. Fewer than three markers

The app needs at least three markers. With fewer, the tablet will not recognize them at all.

4. One marker hiding another

If a marker blocks the view of another, capture the blocking marker fully, remove it, then capture the rest. Do not loosen or move the other markers while you do this.

5. Starting too close

Autofocus needs room. Start about 8 inches away and move in slowly to 5 to 3 inches, with the small cross centered on a marker.

6. The wrong replacement number

Every marker has its own code. A lost AT-01 must be replaced with another AT-01, not the next number up.

More T-Marker guides

Order T-Marker from Dentcore

AOT, URIS and Straumann® kits from $3,000. Not sure which connection you have? Call 844-292-8023 and we will check your MUA before you order.

Shop T-Marker kitsSee how T-Marker works

All T-Marker kits at Dentcore →

For reference only. Specifications, pricing, and availability are as of the publication date and may change without notice. Dentcore is not responsible for discrepancies, always verify with the manufacturer before purchasing. T-Marker™ is a trademark of TruAbutment Inc. Straumann®, Apple®, iPad® and iPhone® are trademarks of their owners.